Features:

The TD-HC-07 hematocrit centrifuge with the TD-DC-071 hematocrit rotor is used for determination of volume fractions of erythrocytes in blood for diagnostic purposes and doping tests. Also it is used to derive plasma for the photometric determination of the bilirubin content of the blood of neonates.
The hematocrit centrifuge with the TD-DC-072 tube rotor is suitable for 1.5/2 mL tubes, which can be used for extraction of DNA and RNA

Categoría:

Descripción

Quick Details
Classification: Laboratory Centrifuge
Brand Name: Four E’S science
Model Number: TD-HC-07 Hematoctrit Centrifuge
Hematoctrit Centrifuge Speed range: 200-14000 increment:10
Hematoctrit Centrifuge Max RCF: 18620 increment 10
Hematoctrit Centrifuge Rotor Capacity: tube rotor :1.5/2ml/X24 (14000rpm) Cappillary tube X24 (12000rpm)
Hematoctrit Centrifuge Time range: Timing operation 30s-99m /Continuous operation
Driving motor: Brushiess DC Motor
Weight: 10KG
Dimension: 280*364*266mm
Power: 220V-240V/110V-120V ,50HZ/60HZ,5A
Acceleration /bracking time: 20S/20S

 

Model
TD-HC-07
Max. Speed [rpm]
200-14000rpm, increment:10
Max. RCF [×g]
18620×g, increment: 10×g
Speed Accuracy [rpm]
±30rpm
Rotor Capacity
Tube rotor: 1.5/2ml/×24(14000rpm); Hematocrit rotor: capillary tubers x 24 (12000rpm)
Run Time
30sec-99min-HOLD (Continuous operation)
Driving Motor
Brushless DC motor
Safety Devices
Door interlock, Over-speed detector; Over-temperature detector;
Automatic internal diagnosis; Automatic rotor identification
Power
Single-phase, 220V-240V, 50HZ/60HZ, 5A;110V-120V, 50HZ/60HZ, 5A
Noise level [dB]
≤60
Acceleration/Braking time
20s↑20s↓
Dimensions [mm]
(L)280×(W)364×(H)266
Weight
10kg
Additional Features
Speed/RCF switch; Short-time run function; sound-alert function