Descripción
Features:
- Unique optical system, 1200 lines/mm grating and imported receiver to ensure its sound performance.
- The optional PC software expands the capabilities of the instrument includes Photometric analysis, Quantitative analysis, Spectral scanning, Kinetics, DNA/protein analysis, Multi-wavelength test, etc.
- Adopts 51 microcomputer system, and store 200 test data and 100 standard curves;
- Plug type deuterium lamp and tungsten lamp transfer lamps without optics debugging.
- Large sample chamber can accommodate 5-100mm cuvettes of all kinds.
Model | K/V1700 | K/V1710 |
Optical System | Single Beam, Grating 1200 lines/mm | |
Wavelength Range | 330-1100nm | 340-1050nm |
Spectral Bandwidth | 2 nm | 4nm |
Wavelength Accuracy | ±0.5 nm | 1nm |
Wavelength Repeatability | <=0.1 nm | 0.2 nm |
Photometric Accuracy | 0.3%T | 0.5%T |
Photometric Repeatability | 0.2%T | |
Photometric Range | 0 – 200%T, – 0.3 – 3A | |
Stray light | 0.05%T | 0.15%T |
Stability | ±0.0001A/h @500nm | ±0.0002A/h @500nm |
Baseline Flatness | ±0.002 A | |
Noise | ±0.001 A | |
Wavelength Setting | AUTO | |
Light Source | Tungsten Halogen Lamp ( 12v,20w ) | |
Display | 128*64 LCD | |
Detector | Silicon Photodiode | |
Output | USB port & Parallel port ( Printer) | |
Power | AC 220V/50Hz; AC 110V/60Hz | |
Dimension | 590*460*220mm | |
Net Weight | 16kg | |
Shipping Size | 700*580*340mm | |
Weight | 20kg |